General Purpose Power Supply (GPDPS) 6 Device PowerSupply
System Reference, January 2001
139
GPDPS Specifications.
The following supply voltage/current range specifications
apply for the General Purpose Device Power Supply:
Table 26 General Purpose DPS: Voltage/Current Range Specifications
Voltage and Current Measurements
For current measurements, four ranges are available. In
test functions that use the GPDPS as a measuring unit,
autoranging is used to find the best current range.
The lowest current range provides the highest measure-
ment accuracy. Note, that this measurement accuracy can
only be guaranteed for a load capacitance less than 1µF
and at least 500 samples. A load capacitance of 1 µF is
only possible if the GPDPS operates within the perfor-
mance range 1. In all other performance ranges the
minimum capacitive load to ensure stable GPDPS opera-
tion is larger. See also “Setting up Performance Ranges”
on page 141.
Mode Range Resolution Accuracy Comments
Voltage force 7...+7 V
–8...+8 V
1mV
1mV
±5mV ±0.1%
±5mV ±0.1%
I
max
= +8 A, –4 A
I
max
= ±4A
Voltage measure 8...+8 V 1 mV ±5mV ±0.1%
Current force
(clamp)
–4A, +8A 1mA ±20 mA ±0.5% 16 samples
Current measure –8...+8 A
–0.3...+0.3 A
–10...+10 mA
–100...+100µA
1mA
30 µA
1 µA
10 nA
±20 mA ±0.1%
±450 µA ±0.15%
±10 µA ±0.1%
±100 nA±0.1%
16 samples
16 samples
C
load
< 100µF,
32 samples
C
load
< 1 µF,
500 samples