Test System Configuration 4 Test Head Filling and DUT Board Considerations
System Reference, January 2001
81
Analog Fill Order
The order of analog filling depends on the test head config-
uration. Therefore, the fill order for the digital dominant
configuration is described in part one of this section while
you can find that one for the analog dominant configura-
tion in part two.
1. The fill order of the analog instruments (see Table 5) is
illustrated in Figure 40 for the digital dominant configu-
ration.
Figure 40 Fill Order of Analog Instruments within the Digital Dominant
Configuration
m
In Figure 40 the six lightly shaded (yellow) boxes are the
half groups of those three groups which can support
analog instrument within the 1024 pin digital dominant
configuration. Thus, for this configuration the analog
instruments can only be placed in those card cages corre-
sponding to groups 6, 7 and 8. In the 512 pin digital domi-
nant configuration the analog instruments can only be
placed in the card cage slots corresponding to group 8.
The analog time analysing or measuring instruments are
placed in the half groups titled MEAS. The analog signal
generators are placed in the half groups titled SOURCE.
Depending on the half group the abbreviations of the
analog instruments correspond either to source or
measurement in the table in Table 5. The numbers next to
an abbreviation assign the count number of an instrument
type. These count numbers increase with increasing
filling of the cages.