6 Device PowerSupply High Current Power Supply (HCDPS)
176 System Reference, January 2001
HCDPS Specifications
Note that in contrast to the General Purpose and High
Voltage DPS, there is only one voltage and current range.
When looking at minimum currents to be measured,
please take into account:
• The base accuracy of the GDPS is about 0.25% of full
scale. This represents 20mA for 8A. Compared to this,
the accuracy of the HCDPS is 200mA for 100A.
• A static load with negative currents is not possible,
hence, no such calibration can be performed.
• The voltage is programmable down to 0V for testing the
protection diodes of the DUT (Continuity Test).
• The current measurements in ganged mode sum up the
individual values of 200mA accuracy for each HCDPS
module to a total value of 400mA.
HCDPS Decoupling Recommendations
This is described under “Decoupling Recommendations”
on page 154.
Type Value Resolution Accuracy
Voltage force 0.3...2.5V (100 A
max
)
2.5...4V (50 A
max
)
1mV ±5mV
Voltage measure 0V...4V 1mV ±5mV
Current clamp +1...100A 100mA for I≤10A: ±(200mA+0.5% of value)
for I>10A: ±(200mA+2% of value)
Current measure ±100A 10mA for 0-100A: ±(100mA+0.4% of value)
negative currents without calibra-
tion
Table 33 High Current DPS: Voltage/Current Range Specifications